Indirect conversion
Scintillators make light.
A scintillator absorbs the X-ray or gamma photon and re-emits its energy as a short flash of visible light. A photodetector - photomultiplier, SiPM or silicon photodiode - turns that light into an electrical pulse. Two steps, each with its own losses.
photon → electrons & holes in the crystal → activator (Tl⁺, Ce³⁺) → light → photodetector
Physics
From a photon to a flash.
Stopping power
Density and Z decide how thick the crystal must be (BGO, LYSO, GOS are the densest).
Light yield
Photons per MeV: more light → more photo-electrons → better energy resolution.
Decay time
How fast the light comes out: sets count rate, pile-up and PET timing.
Emission spectrum
Must match the photodetector: blue for PMTs, green for silicon photodiodes.
Lab 1
Absorption of the scintillators.
Lab 2
Bright or fast?
Lab 3
Match the light to the photodetector.
Photomultiplier (PMT)
Photocathode + dynode chain, gain ~10⁶, very low noise, ns timing. Bulky, ~1 kV, magnetic-field sensitive. Standard for NaI spectrometers.
SiPM
Array of Geiger-mode avalanche micro-cells, gain ~10⁶ at 30-60 V. Compact and MRI-compatible - used in modern TOF-PET. Dark counts and crosstalk.
Silicon photodiode
No gain but QE up to 80-90 % in the green. Cheap and stable - used under CsI(Tl) in X-ray panels and handheld spectrometers.
Lab 4
Silicon alone vs scintillator + silicon.
Lab 5
Thickness and resolution.
Data
Scintillator properties.
| Scintillator | Formula | ρ g/cm³ | Z eff | Light yield /MeV | Decay | Emission | Notes |
|---|---|---|---|---|---|---|---|
| NaI(Tl) | NaI | 3.67 | 49.7 | 38,000 | 230 ns | 415 nm | Classic gamma spectroscopy, SPECT cameras; hygroscopic. |
| CsI(Tl) | CsI | 4.51 | 54 | 54,000 | 1 µs | 550 nm | Bright, green light ideal for Si photodiodes; X-ray flat panels. |
| LaBr₃(Ce) | LaBr3 | 5.08 | 45.2 | 63,000 | 16 ns | 380 nm | Best-resolution scintillator (~3 % at 662 keV); fast. |
| BGO | Bi4Ge3O12 | 7.13 | 71.5 | 8,200 | 300 ns | 480 nm | Very dense, high Z, but dim. PET, Compton suppression. |
| LYSO(Ce) | Lu1.8Y0.2SiO5 | 7.1 | 62.3 | 30,000 | 40 ns | 420 nm | Dense and fast: the standard for time-of-flight PET. |
| GOS (Gd₂O₂S:Tb) | Gd2O2S | 7.34 | 59.4 | 60,000 | 600 µs | 545 nm | Ceramic/powder screens for X-ray and CT detectors. |
| Plastic (PVT) | C9H10 | 1.03 | 5.7 | 10,000 | 2.4 ns | 425 nm | Cheap, large, fast - but low Z: almost no photopeak. |
Compare with direct conversion