Gamma-ray
spectroscopy.

Five steps from a single gamma photon to the activity of a sample - how a detector turns a flash in a crystal into a spectrum. Everything here is live and computed in your browser.

loading interactive lab…
  1. 01

    Photon in the crystal

    Photoeffect, Compton, pair production - and where each event lands in the spectrum.

  2. 02

    Geometry & efficiency

    Why crystal size and distance decide how many photons end up in the peak.

  3. 03

    Materials & resolution

    NaI, LaBr₃, CZT, CsPbBr₃ perovskite, HPGe: why some peaks are sharper.

  4. 04

    From deposit to signal

    Light or charge, trapping and μτ, and how a pulse becomes one count in one channel.

  5. 05

    Reading the spectrum

    Peaks, edges, escape and sum peaks - and from counts to becquerels.

Step 1

What a photon does inside the crystal

A gamma ray enters the crystal like a ball thrown into a dense forest. It can stop at once and give up all its energy, bounce off a "tree" and carry part of the energy away, or fly straight through. The instrument only sees what the photon left behind in the crystal. Fire photons and watch the spectrum grow.
loading interactive lab…

Full-energy peak

The photon gave all its energy. The peak position identifies the nuclide; its area tells how much of it there is.

Compton plateau

The photon bounced and escaped, leaving part of its energy. The plateau ends sharply at the Compton edge.

Satellite bumps & peaks

Backscatter and lead X-rays come from the surroundings; escape peaks come from the crystal itself. No new nuclide behind them.

Every outcome, one at a time

Pick an outcome - the scene replays one typical history in slow motion and marks where it lands in the spectrum. Each outcome uses the nuclide where it is easiest to see.

loading interactive lab…
E′ = E / (1 + (E / mₑc²)(1 − cos θ))  ·  Compton edge = E·2ε/(1+2ε), ε = E/mₑc²  ·  backscatter ≈ E/(1+2ε)
E = incident photon energy, θ = scattering angle, mₑc² = 511 keV

Step 2

Geometry: why size and distance matter

The source shines in all directions; the crystal catches only what flies into it - like a bucket in the rain. A smaller or more distant bucket catches fewer drops. The full-energy-peak efficiency says which fraction of emitted photons ends up in the peak.
loading interactive lab…

One curve per setup

The same sample in a different container, at a different gap, or with a different density gives a different curve. That is why every measurement geometry needs its own efficiency calibration (measured with reference sources or computed with Monte-Carlo, like here).

Step 3

Detector materials and energy resolution

Stopping the photon is half the job; measuring its energy precisely is the other half. The energy resolution (peak width, FWHM) decides whether close lines can be separated - for example Co-60's 1173 and 1332 keV, or nuclides in a mixed sample. Acquire a spectrum and compare two detectors.
loading interactive lab…
loading interactive lab…
loading interactive lab…
MaterialTypeDensity g/cm³Z effFWHM @ 662 keVSignal carriersSpeedOperationTypical use
NaI(Tl)Scintillator3.67507%≈38 000 photons/MeV230 nsRoom T (hygroscopic)Survey spectrometers, SPECT gamma cameras
CsI(Tl)Scintillator4.51546.5%≈54 000 photons/MeV≈1 µsRoom TFlat-panel X-ray, handheld spectrometers
LaBr₃(Ce)Scintillator5.08472.9%≈63 000 photons/MeV16 nsRoom T (hygroscopic)High-resolution scintillation spectroscopy
BGOScintillator7.137510.5%≈8 200 photons/MeV300 nsRoom TPET (older systems), Compton suppression
LYSO(Ce)Scintillator7.1669%≈30 000 photons/MeV≈40 nsRoom TTime-of-flight PET
CdZnTe (CZT)Semiconductor5.78492%W ≈ 4.6 eV per e–h pairdrift ≈ 1 µsRoom TPhoton-counting CT, cardiac SPECT, isotope ID
CsPbBr₃ perovskiteSemiconductor4.55651.4%W ≈ 5.3 eV per e–h pairdrift ≈ µsRoom T (2-70 °C shown)Emerging: melt-grown room-temperature spectroscopy, photon-counting X-ray
FAPbBr₃ perovskiteSemiconductor3.8611.7%W ≈ 6–7 eV per e–h pair (est.)drift ≈ µsRoom T (to >130 °C shown)Emerging: solution-grown gamma spectroscopy & X-ray photon counting
HPGeSemiconductor5.32320.2%W ≈ 2.96 eV per e–h pairdrift ≈ 0.5 µsCooled to ~77 KLaboratory gold standard for gamma spectroscopy

Typical literature values; real numbers depend on crystal size, quality and electronics. Perovskite values (CsPbBr₃, FAPbBr₃) are from recent research devices; champions reach 0.8 % and 0.7 % at 662 keV.

Why semiconductors give sharper peaks

One electron-hole pair costs only ~3-6 eV, while a scintillator + PMT needs ~100 eV or more per photo-electron. More information carriers → smaller relative fluctuation (∝ 1/√N) → narrower peaks. HPGe must be cooled to ~77 K; CZT and lead-halide perovskites such as CsPbBr₃ work at room temperature.

Step 4

From deposited energy to a signal

A scintillator converts the energy into light, a semiconductor into electric charge. Either way the result is a tiny current pulse whose size is proportional to the energy. In semiconductors, carriers can be trapped before they reach the electrodes - the key material parameter is μτ (mobility × lifetime).
loading interactive lab…
CCE(z) = (λₑ/d)·(1 − e^(−(d−z)/λₑ)) + (λₕ/d)·(1 − e^(−z/λₕ))
Hecht equation, planar detector of thickness d, interaction at depth z from the cathode, λ = μτ·V/d

Pulse → channel

The preamplifier integrates the charge into a step; the shaping amplifier turns it into a smooth pulse whose height is measured by an ADC. The multichannel analyser (MCA) adds one count to the channel of that height. Millions of such counts make the spectrum.

loading interactive lab…

Step 5

Reading the spectrum - and getting becquerels

Each feature of a spectrum has a physical origin. Click the features to see what produces them, then turn a peak area into an activity with its uncertainty and detection limit.
loading interactive lab…
loading interactive lab…

Error shrinks as √t

Measuring 4× longer gives a 2× smaller uncertainty.

Detection limit

Below it the instrument cannot tell the sample from background - the honest answer is "less than X Bq" (ISO 11929, Currie).

Bq/kg

Activity divided by mass: weigh the sample and fill the container exactly as in the calibration geometry.

Summary

The whole chain in one line.

Photon interacts→Energy deposited→Light / charge→Pulse→MCA channel→Spectrum→Peak area→Bq

The same detectors - in arrays of thousands - form X-ray, CT, PET and SPECT scanners. See how in the imaging section.

Models on this page are simplified and run in your browser: Monte-Carlo photon transport in a NaI(Tl) cylinder (photoeffect with iodine K X-ray escape, Klein-Nishina Compton scattering, pair production), smooth fits to NIST XCOM cross-sections, and typical literature resolutions. Numbers are approximate.