Single photon emission CT
Seeing single gamma rays.
SPECT tracers such as Tc-99m emit one 140 keV gamma per decay - there is no partner photon to tell us the direction. A lead collimator in front of the camera solves this by letting through only photons travelling along its holes.
Tc-99m
140 keV
half-life 6 h, most used isotope
Collimator efficiency
≈ 10⁻⁴
1 in ~10 000 photons passes
System resolution
8-12 mm
at 10 cm distance
Views
60-128
over 180° or 360°
Interactive
SPECT lab: collimators and rotation
A cardiac phantom with a perfusion defect and a hot spot. The camera rotates through 64 views. Wider collimator holes accept more photons (less noise) but each hole "sees" a wider cone - the image gets blurrier with depth. Green photons pass through the holes; red ones are stopped by the lead septa.
Things to try
- Set acceptance to 0.5° (high-resolution): sharp but noisy. 5° (high-sensitivity): smooth but blurred.
- Cut time per view to 2 s: can you still see the defect in the heart wall?
Hardware
The Anger camera and beyond
One large NaI(Tl) crystal (≈9.5 mm thick) viewed by an array of PMTs. The position of each flash is computed from the light-sharing between PMTs; the total light gives the energy, used to reject scattered photons (window around 140 keV).
Pixelated CdZnTe modules measure position directly with ~2.5 mm pixels and ~5-6 % energy resolution at 140 keV (vs ~9-10 % for NaI). Dedicated cardiac and whole-body CZT systems give better contrast, dual-isotope imaging and lower dose.
SPECT is often combined with CT (SPECT/CT): the CT gives anatomy and an attenuation map to correct the SPECT data.