Single photon emission CT

Seeing single gamma rays.

SPECT tracers such as Tc-99m emit one 140 keV gamma per decay - there is no partner photon to tell us the direction. A lead collimator in front of the camera solves this by letting through only photons travelling along its holes.

loading interactive lab…

Tc-99m

140 keV

half-life 6 h, most used isotope

Collimator efficiency

≈ 10⁻⁴

1 in ~10 000 photons passes

System resolution

8-12 mm

at 10 cm distance

Views

60-128

over 180° or 360°

Interactive

SPECT lab: collimators and rotation

A cardiac phantom with a perfusion defect and a hot spot. The camera rotates through 64 views. Wider collimator holes accept more photons (less noise) but each hole "sees" a wider cone - the image gets blurrier with depth. Green photons pass through the holes; red ones are stopped by the lead septa.

loading interactive lab…

Things to try

  • Set acceptance to 0.5° (high-resolution): sharp but noisy. 5° (high-sensitivity): smooth but blurred.
  • Cut time per view to 2 s: can you still see the defect in the heart wall?

Hardware

The Anger camera and beyond

Anger camera (1957 → today)

One large NaI(Tl) crystal (≈9.5 mm thick) viewed by an array of PMTs. The position of each flash is computed from the light-sharing between PMTs; the total light gives the energy, used to reject scattered photons (window around 140 keV).

CZT cameras

Pixelated CdZnTe modules measure position directly with ~2.5 mm pixels and ~5-6 % energy resolution at 140 keV (vs ~9-10 % for NaI). Dedicated cardiac and whole-body CZT systems give better contrast, dual-isotope imaging and lower dose.

SPECT is often combined with CT (SPECT/CT): the CT gives anatomy and an attenuation map to correct the SPECT data.