Computed tomography
Many shadows → one slice.
A CT scanner takes hundreds of X-ray projections while the tube and detector spin around the patient (≈ 0.25-0.5 s per turn). A computer then solves the inverse problem: which map of attenuation coefficients explains all these projections?
Rotation
0.25 s
fastest clinical gantries
Detector rows
64-320
up to 16 cm coverage
Views / turn
≈ 1000-2000
projections
Units
HU
water 0, air −1000, bone +700…+3000
Interactive
CT lab: from sinogram to slice
Each row of the sinogram is one projection (one tube angle). A point in the object traces a sine curve - hence the name. Simple backprojection smears every projection back across the image and produces a blurry result; applying a ramp filter first (filtered backprojection, FBP) restores sharp edges.
Things to try
- Switch to No filter: the image is blurred by a 1/r point-spread function.
- Use only 12 or 30 views: streak artifacts appear (angular undersampling).
- Lower the dose: noise grows; the Shepp-Logan filter trades a little sharpness for less noise.
Theory
The maths in two lines
Modern scanners use iterative and deep-learning reconstruction to cut dose, and photon-counting CT (CdTe/Si detectors, clinical since 2021) adds energy information: material maps (iodine, calcium), virtual monoenergetic images and higher spatial resolution.
Next topic
Positron emission tomography (PET)