Computed tomography

Many shadows → one slice.

A CT scanner takes hundreds of X-ray projections while the tube and detector spin around the patient (≈ 0.25-0.5 s per turn). A computer then solves the inverse problem: which map of attenuation coefficients explains all these projections?

loading interactive lab…

Rotation

0.25 s

fastest clinical gantries

Detector rows

64-320

up to 16 cm coverage

Views / turn

≈ 1000-2000

projections

Units

HU

water 0, air −1000, bone +700…+3000

Interactive

CT lab: from sinogram to slice

Each row of the sinogram is one projection (one tube angle). A point in the object traces a sine curve - hence the name. Simple backprojection smears every projection back across the image and produces a blurry result; applying a ramp filter first (filtered backprojection, FBP) restores sharp edges.

loading interactive lab…

Things to try

  • Switch to No filter: the image is blurred by a 1/r point-spread function.
  • Use only 12 or 30 views: streak artifacts appear (angular undersampling).
  • Lower the dose: noise grows; the Shepp-Logan filter trades a little sharpness for less noise.

Theory

The maths in two lines

p(θ, t) = ln(I₀ / I) = ∫ μ(x, y) ds
Radon transform: p(θ, t) is the line integral of μ along the ray at angle θ and offset t
μ(x, y) = ∫₀^π (p(θ, ·) * h)(x cos θ + y sin θ) dθ
Filtered backprojection: h is the ramp (|ω|) filter

Modern scanners use iterative and deep-learning reconstruction to cut dose, and photon-counting CT (CdTe/Si detectors, clinical since 2021) adds energy information: material maps (iodine, calcium), virtual monoenergetic images and higher spatial resolution.

Next topic