Positron emission tomography

Two photons, back to back.

A PET tracer such as ¹⁸F-FDG emits positrons. Each positron travels ~1 mm, meets an electron and annihilates into two 511 keV photons flying in opposite directions. Detecting both at the same time defines a line on which the decay happened.

loading interactive lab…

Photon energy

511 keV

always the same - by E = mc²

Coincidence window

≈ 4 ns

both photons must arrive within it

TOF resolution

200-400 ps

≈ 3-6 cm along the line

Crystals

LYSO / BGO

dense, fast, high Z

Interactive

PET lab: lines of response

The phantom has a heart wall, a small tumour and a bladder. Every detected pair draws a line of response (LOR); backprojecting many LORs reveals the activity. Turn on scatter and randoms to see what blurs real images, and time-of-flight to see how measuring the arrival-time difference localises each event along its line.

loading interactive lab…

Why time-of-flight helps

If one photon arrives Δt earlier, the annihilation happened c·Δt/2 closer to that detector. With 200 ps timing that is ±3 cm - not a pixel, but enough to cut noise roughly by √(D/Δx) for a patient of diameter D.

Physics

What limits PET image quality

Δx = c · Δt / 2
Δx: position uncertainty along the LOR, c: speed of light, Δt: coincidence timing resolution (FWHM)
  • Positron range (≈0.6 mm for ¹⁸F, several mm for ⁶⁸Ga, ⁸²Rb) and non-collinearity (photons are not exactly 180° apart, ±0.25°) blur the image.
  • Depth of interaction: oblique photons penetrate several crystals - parallax error at the edge of the field.
  • Scatter is rejected with an energy window around 511 keV - so energy resolution matters.
  • Randoms grow with activity² and shrink with a narrower coincidence window - so timing matters.