Positron emission tomography
Two photons, back to back.
A PET tracer such as ¹⁸F-FDG emits positrons. Each positron travels ~1 mm, meets an electron and annihilates into two 511 keV photons flying in opposite directions. Detecting both at the same time defines a line on which the decay happened.
Photon energy
511 keV
always the same - by E = mc²
Coincidence window
≈ 4 ns
both photons must arrive within it
TOF resolution
200-400 ps
≈ 3-6 cm along the line
Crystals
LYSO / BGO
dense, fast, high Z
Interactive
PET lab: lines of response
The phantom has a heart wall, a small tumour and a bladder. Every detected pair draws a line of response (LOR); backprojecting many LORs reveals the activity. Turn on scatter and randoms to see what blurs real images, and time-of-flight to see how measuring the arrival-time difference localises each event along its line.
Why time-of-flight helps
Physics
What limits PET image quality
- Positron range (≈0.6 mm for ¹⁸F, several mm for ⁶⁸Ga, ⁸²Rb) and non-collinearity (photons are not exactly 180° apart, ±0.25°) blur the image.
- Depth of interaction: oblique photons penetrate several crystals - parallax error at the edge of the field.
- Scatter is rejected with an energy window around 511 keV - so energy resolution matters.
- Randoms grow with activity² and shrink with a narrower coincidence window - so timing matters.