Projection radiography
An X-ray image is a map of shadows.
X-rays from a tube pass through the body. Bone, soft tissue and contrast agents absorb different fractions, and the detector records what is left. Contrast, noise and dose are all controlled by the photon energy spectrum and by the detector.
Tube voltage
40-150 kVp
sets the maximum photon energy
Bone vs tissue
≈ 3-6×
μ ratio at 30-60 keV
Iodine K-edge
33.2 keV
why iodine is a great contrast agent
Chest X-ray dose
≈ 0.02 mSv
≈ 3 days of natural background
Physics
Beer-Lambert: how much gets through
Because μ depends strongly on energy (photoeffect ∝ Z⁴/E³), low-energy photons give the best contrast but are absorbed in the patient and add dose without reaching the detector. Raising kVp lowers dose and noise but washes out contrast - every exam is a compromise.
Interactive
X-ray lab
The phantom is a limb with two bones, an iodine-filled vessel and a faint soft-tissue nodule. Change the beam and the detector and watch the image and the contrast-to-noise ratios.
Things to try
- Lower the mAs until the soft nodule disappears in noise: that is the dose-image quality trade-off.
- Move kVp from 50 to 120 and watch bone and iodine contrast fall. The iodine vessel is brightest when many photons sit just above 33 keV.
- Switch to photon counting: equal weighting of low-energy photons and no electronic noise raise CNR at the same dose.
Hardware
Digital X-ray detectors
Needle-structured CsI(Tl) or Gd₂O₂S scintillator on an amorphous-silicon photodiode/TFT array. The workhorse of radiography, fluoroscopy and cone-beam CT.
Amorphous selenium converts X-rays directly into charge - very sharp images, used in mammography. Limited stopping power at higher energies.
CdTe/CZT (and emerging perovskite) sensors bump-bonded to counting ASICs. Each photon is counted and energy-binned: no electronic noise, spectral imaging, smaller pixels.